Pular para o conteúdo principal
GE Healthcare · MRExcite225230450334

2252304 Multi Coil Bias Fault Due to Driver Module for Excite.

Symptom

Unable to Scan With Multi Coil Bias Faults Unable to scan with Multi Coil Bias Faults

Error message

2252304 Multi coil Bias Fault Receive Fault. Wed May 31 11:05:41 2006 Error: 2252304 Host: PHDGE3TMR Process: NSP File: SCP:DCB Line: 1143 Exception Class: Unknown Severity: Unknown Function: RF: RF Bias PSD: Spin Echo Coil: XXXXX Scan: 50334/1 Multicoil Bias Fault - Receive Fault. Try re-seating all coil connectors. MCD2 Channel 4 of 16, counting from 1 Value: 125 mA, Limit: 98 mA Mask of failed channels: 0xd8 MCD1 1-8 : A or C MCD1 9-16 : B MCD2 1-8 : B or Legacy MC MCD2 9-16 : C->

Applies to

Models
Excite
Product
Excite HDx Includes "Excite HDi" system. 14.0 Software only.
Software
14.0 M5 Excite HDx HDi Platforms only.
Component
SecondOpinion Excite HD & HDx Driver Module 5110612-2 See comments in resolution concept for second opinion requirements before ordering this part
Area
MR

Full cause & step-by-step resolution

The complete cause analysis, the step-by-step remedy and the attached service documents for this GE Healthcare MR issue are available to FSELIB subscribers — re-authored, web-first, searchable.

7-day free trial · no credit card required

Related GE Healthcare MR issues