GE Healthcare · MRExciteSigna22185372252293225231150422
Unable to Scan with Driver Module Errors due to CRM RF Coil Short Circuit.Scan Stopped with Driver Module Errors and a F…
Unable to Scan with Driver Module Errors due to CRM RF Coil Short Circuit.Scan Stopped with Driver Module Errors and a Foul Smell in the Scan Room.Replace Shorted CRM RF Coil.Excite HD
Symptom
Scan Stopped with Driver Module Errors and a Foul Smell in the Scan Room. Scan stopped suddenly and driver module error occurred . There is foul smell in the scan room.
Error message
2218537 Scan stopped, Scan Stopped due to error. Driver Module error occurred.. 2218537 HCMR scn MsaScanMgr.m 3038 Scan stopped, Scan Stopped due to error : Driver Module error occurred.. 2252293 The Driver Module Has Detected a Fault. Wed Apr 21 20:00:24 2004 Error: 2252293 Host: lx-mr Process: NSP File: SCP:DCB Line: 1986 The Driver Module Has Detected a Fault.^A Details of the Fault can be found in other error messages.-> 2252311 DD Driver Fault Detected - Short Circuit. Sun Mar 2 15:54:06 2008 Error: 2252311 Host: MR3T Process: NSP File: SCP:DCB Line: 1846 Exception Class: Unknown Severity: Unknown Function: RF: RF Bias PSD: Spin Echo Coil: HEAD Scan: 50422/2 DD Driver Fault Detected - Short Circuit. Source of Fault is BD2 (Dynamic Disable Boards at Front of Body Coil). Value: 50 mA Limit: 40 mA->
Applies to
- Models
- Excite, Signa
- Product
- Excite HD For 12.0 1.5T Systems only.
- Software
- 12.0 Excite HD and 3T Excite HD products only
- Component
- 2208999 CRM RF COIL AND CASE. 2208999 CRM RF COIL AND CASE.
- Area
- MR
Full cause & step-by-step resolution
The complete cause analysis, the step-by-step remedy and the attached service documents for this GE Healthcare MR issue are available to FSELIB subscribers — re-authored, web-first, searchable.
7-day free trial · no credit card required
Related GE Healthcare MR issues
- Running grafidy3, and find there are spikes on the waveforms in the manual prescan display, both windows.
- TPS/MGD fails to download or reset. mgd_term windows fail to mount file system.
- Fails to Scan with TORSO Coil.
- The Carriage Fails to Advance to Latch to the Table and to Retract When it is Raised, Undocked or Lowered.
- AE title has changed on destination host.