Unable to Scan with Driver Module Errors due to CRM RF Coil Short Circuit.Scan Stopped with Driver Module Errors and a F…

Unable to Scan with Driver Module Errors due to CRM RF Coil Short Circuit.Scan Stopped with Driver Module Errors and a Foul Smell in the Scan Room.Replace Shorted CRM RF Coil.Excite HD

Symptom

Scan Stopped with Driver Module Errors and a Foul Smell in the Scan Room. Scan stopped suddenly and driver module error occurred . There is foul smell in the scan room.

Error message

2218537 Scan stopped, Scan Stopped due to error. Driver Module error occurred.. 2218537 HCMR scn MsaScanMgr.m 3038 Scan stopped, Scan Stopped due to error : Driver Module error occurred..

2252293 The Driver Module Has Detected a Fault. Wed Apr 21 20:00:24 2004 Error: 2252293 Host: lx-mr Process: NSP File: SCP:DCB Line: 1986 The Driver Module Has Detected a Fault.^A Details of the Fault can be found in other error messages.->

2252311 DD Driver Fault Detected - Short Circuit. Sun Mar 2 15:54:06 2008 Error: 2252311 Host: MR3T Process: NSP File: SCP:DCB Line: 1846 Exception Class: Unknown Severity: Unknown Function: RF: RF Bias PSD: Spin Echo Coil: HEAD Scan: 50422/2 DD Driver Fault Detected - Short Circuit. Source of Fault is BD2 (Dynamic Disable Boards at Front of Body Coil). Value: 50 mA Limit: 40 mA->

Applies to

Models
Excite, Signa
Product
Excite HD For 12.0 1.5T Systems only.
Software
12.0 Excite HD and 3T Excite HD products only
Component
2208999 CRM RF COIL AND CASE. 2208999 CRM RF COIL AND CASE.
Area
MR

Full cause & step-by-step resolution

The complete cause analysis, the step-by-step remedy and the attached service documents for this GE Healthcare MR issue are available to FSELIB subscribers — re-authored, web-first, searchable.