Pular para o conteúdo principal
GE Healthcare · MRExcite225008322501092245309162784722500822253851

Gradient Data Overrange Errors Due to STIF Board.

Symptom

Will Not Scan with Gradient Errors.

Error message

2250083 Y Axis Gradient Data Overrange Occurred. Error: 2250083 Host: CHCMR2 Process: NSP File: SCP:Grad Line: 3255 Exception Class: Unknown Severity: Unknown Function: Gradient: N/A PSD: FSE-XL Coil: HD TRknee PA Scan: 2458/7 Y Axis Gradient Data Overrange Occurred.-

2250109 The GP is reporting a Gradient Clock Stop. The GP stopped receiving a gradient clock signal from the systems cabinet.

2245309 Gradient Diagnostic test has been aborted because GP detected a bad clock input 2245309 WCLAGEMR2MR diagCoord grad_lib.c 924 Diagnostic: Grad Frame/Clk Iteration: 25 of infinite Action: 1 Task ID: 0x3fc4380 Time: 1627847 Object: SCP Test Task Function: Gradient: N/A Gradient Diagnostic test has been aborted because GP detected a bad clock input.-> Extended error log: "Check placement of Clock and X,Y,Z data fibers between the STIF and GP (look for swapped fibers between clock and x/y/z data fibers)" 2245309 WCLAGEMR2MR diagCoord grad_lib.c 924 Diagnostic: Grad Frame/Clk Iteration: 25 of infinite Action: 1 Task ID: 0x3fc4380 Time: 1627847 Object: SCP Test Task Function: Gradient: N/A Gradient Diagnostic test has been aborted because GP detected a bad clock input.-> Extended error log: 

Applies to

Models
Excite
Product
Excite HDe. For Hino HDe 14.0 and HDe 15.0 Software Systems only
Software
HDe 14.0 For Hino Excite HDe 14.0 systems only.
Component
2280952-3 STIF Board See Long Description SRF TRF I/O or STIF 2280952 crosses up in FEMC to 2280952-3. Part Description is "STIF Assembly".
Area
MR

Full cause & step-by-step resolution

The complete cause analysis, the step-by-step remedy for this GE Healthcare MR issue are available to FSELIB subscribers — re-authored, web-first, searchable.

7-day free trial · no credit card required