GE Healthcare · MRExciteSigna22481302248122

PSD download errors on any 8 channel coil Due to DRF Board.

Symptom

Unable to Scan With The Download of a Prescribed Series Failed Message

Error message

2248130 APS DC Setup DRF Channels Error. DRFs can only support -x channels 2248130 Host: UMIMR3 Process: NSP File: APS:DacqControl::processPkt Line: 513 APS DC setup DRF channels error DRFs can only support -1 channels Software error, please perform a TPS Reset

2248122 Number of Receiver Prescribed is Over the DRF Capability. 2248122 Host: UMIMR3 Process: NSP File: APS:DrfDriver::setupChannel Line: 20 Number of receiver prescribed is over the DRF capability. Maximum receivers supported at this system is 7 Please change prescription.

Applies to

Models
Excite, Signa
Product
Excite Includes Excite 1 and Excite 2, 10.0, 11.0 and 11.1 Software.
Software
11.0 Only for 1.5T Excite product. 12.0 is Excite HD and 14.0 is Excite HDx. Note, 3T Excite goes with software G3, not 11.0 or 11.1.
Component
2215142 DRF board
Area
MR