GE Healthcare · MRExcite
Modality Work list Schedule window will not open Due to WL Files.
Symptom
Modality Work list Schedule window will not open following a power issue at the site Site experienced a power issue where the /usr/g/hisris/WLFilterFile and/or the /w/config/WLSystem.cfg files were corrupted. The issues with these files were corrected remotely, but the techs could not confirm the fix because they were scanning at the time. Ultimately, when they were able to test the fix, the Schedule window did not open.
Error message
Error code 4 WLnewPatientId Could Not Get valid Worklist Error seen in /usr/g/hisris/wlsErrors file: Wed 02-Feb-05 06:26:48 WLscanCallbacks.c (1710) Error code: 4 WLnewPatientId(): Could Not Get valid Worklist Error seen in /usr/g/hisris/wlsErrors file: cd /usr/g/hisris more wlsErrors Wed 02-Feb-05 06:26:48 WLscanCallbacks.c (1710) Error code: 4 WLnewPatientId(): Could Not Get valid Worklist
Applies to
- Models
- Excite
- Product
- Excite Includes Excite 1 and Excite 2, 10.0, 11.0 and 11.1 Software.
- Software
- 11.1 For product 1.5T Excite only. Note, 3T Excite goes with software G3, not 11.0 or 11.1.
- Area
- MR
Full cause & step-by-step resolution
The complete cause analysis, the step-by-step remedy for this GE Healthcare MR issue are available to FSELIB subscribers — re-authored, web-first, searchable.
7-day free trial · no credit card required
Related GE Healthcare MR issues
- TPS/MGD fails to download or reset. mgd_term windows fail to mount file system.
- The Carriage Fails to Advance to Latch to the Table and to Retract When it is Raised, Undocked or Lowered.
- Unable to Load Applications Software during restoreinfo Due to Corrupt saveinfo Disk.
- CWA ID100315. Auto and Manual Prescan Fails.
- Ghosted or Smeared Head Images Using Head Coil due to Head TR Assembly.
- A Copper Coolant Line or Pipe Internal to the MCS Cabinet Chiller Burst due to Water Remaining in Unit Freezing During Shipment.