INX1X1, loosing fluoro during long procedure, tube heat units yellow, but not red. error message on screen "Thermal issu…

INX1X1, loosing fluoro during long procedure, tube heat units yellow, but not red. error message on screen "Thermal issue ; Xray stopped in 60 s.

Symptom

Innova, loosing fluoro during long procedure, tube heat units yellow, but not red. Error message on screen "Thermal issue ; Xray stopped in 60 s. Secure patient and call service"

Error message

3602056 followed by 3602060 3602056: Root Cause: JEDI: Tank temperature too high This warning informs the user that, due to overheating, the exposure will soon be forbidden. It may point to a cooling device problem 3602060: Root Cause: JEDI: Inverter temperature too high Related to a hardware defect

Applies to

Models
Innova
Product
Innova X100 Monoplane Innova X100 Monoplane from product name Innova 2100, 3100 and 4100 Innova X100 Monoplane Innova 2100, 3100 and 4100
Software
Unity 3D M4 SP2 Unity 3D M4 SP2 Fwd Prod. DL Release : DL_UNITY_4.17.5 Ghost DL : 21.22.1 DL : DL_Unity_4.17.5 DL Atlas Core : 4.2-8_5 RTAC Atlas Core : 4.2-8_5 IDC : 5.26 Calypso : 4.37 Calypso Uptime : CLYUPT.UNITY_4.15 Positioner : RR POS LBY 1.4.4U_13 Jedi : P636RC1
Area
X-ray Vascular

Full cause & step-by-step resolution