GE Healthcare · X-ray VascularAdvantxInnova3005030049
No Acquisition exposure, fluoro or record
Symptom
No Acquisition exposure, fluoro or record will not fluoro or record after boot. No issue with Positioner movement, DL, Rtac and detector is up. Innova 2100iq system will not fluoro or record after boot. No issue with Positioner movement, DL, Rtac and detector is up. Error 3200011
Error message
POWER UNIT: KV FAILURE: 42E5: 42E6 Getting error codes 42E5 (Atlas 30050) POWER UNIT : THYRISTOR TQ FAILURE, 42E6 (Atlas 30049) POWER UNIT : KV FAILURE
Applies to
- Models
- Advantx, Innova
- Product
- Advantx E Monoplane Advantx E Systems advantx e, vamp Advantx E generator Systems Advantx E Biplane Advantx E biplane systems: LCLP+ (22 cm II), LCN+ (32 cm II), LCALP+ (40 cm frontal plane, 32 cm lateral plane) Advantx E biplane systems: LCLP+ (22 cm II), LCN+ (32 cm II), LCALP+ (40 cm frontal plane, 32 cm lateral plane) Advantx, LFX Advantx product line LFX Advantx product line
- Component
- Advantx-E VAMP Advantx-E VAMP Advantx-E VAMP
- Area
- X-ray Vascular
Full cause & step-by-step resolution
The complete cause analysis, the step-by-step remedy and the attached service documents for this GE Healthcare X-ray Vascular issue are available to FSELIB subscribers — re-authored, web-first, searchable.
7-day free trial · no credit card required
Related GE Healthcare X-ray Vascular issues
- The Worklist does not refresh on the system when performing a query.
- Innova System boot failure
- INX1X; Message Limit switch engaged; LP4 Lateral Detector lift will not move after going to max SID using the Emergency Backout button on the lift cover because the lift hit maximum (Max) SID limit sw…
- Firewall is broadcasting DHCP addresses
- Cooling Failure; Shut System down and call Service is displayed on the Innova Image monitor
- No gantry motion is reported by the customer.