The 8 Channel Invivo Wrist coil is compatible with HDe 15.0 only.
Symptom
Is the Invivo 8 Channel Wrist Coil compatible with HDe 14.0 See Long Description A shading problem on wrist images exists with a new Invivo 8-channel Wrist coil (part number 5160986-2, M3335LJ). The system is a HDe 14.0 which is using only 4 channels instead of all 8 on this coil. Manual Prescan has only 4 channels displayed during prescan. The coil config file is incorrect because we had to manually enter all the coil parameters due to the Coilconfig.cfg did not contain the 8-channel wrist coil. It only contains the 4-channel coil parameters. Also Healthpage detects an error with the coil config file.
Applies to
- Models
- Excite, Signa
- Product
- Excite HDe. For Hino HDe 14.0 and HDe 15.0 Software Systems only
- Software
- HDe 15.0 For Hino Excite HDe 15.0 systems only.
- Area
- MR
Full cause & step-by-step resolution