GE Healthcare · MRExcite22547822254791
The UPM Calibration files are corrupt and create an unrecoverable condition.
Symptom
Universal Power Monitor faults while calibrating or running calibration scans. Power monitor faults seen during EXCITE HD System Upgrades.
Error message
2254782 The Universal Power Monitor 1 has detected Peak Power Exceeded on the Narrow Band Body Amplifier. Sun Nov 20 12:49:19 2005 Error: 2254782 Host: HDMR Process: NSP File: SCP:UPM Line: 657 The Universal Power Monitor 1 has detected Peak Power Exceeded on the Narrow Band Body Amplifier. Peak value detected = 2236 2254791 The Universal Power Monitor X detected a Pulse Width fault on the Narrow Band Body Amplifier Sun Nov 20 12:43:16 2005 Error: 2254791 Host: HDMR Process: NSP File: SCP:UPM Line: 1231 The Universal Power Monitor 1 detected a Pulse Width fault on the Narrow Band Body Amplifier. Detected Width = 130 uS. Max Width Limit = 0 uS.
Applies to
- Models
- Excite
- Product
- Excite HDx Includes "Excite HDi" system. 14.0 Software only.
- Software
- 14.0 M5 Excite HDx HDi Platforms only.
- Area
- MR
Full cause & step-by-step resolution
The complete cause analysis, the step-by-step remedy for this GE Healthcare MR issue are available to FSELIB subscribers — re-authored, web-first, searchable.
7-day free trial · no credit card required
Related GE Healthcare MR issues
- TPS/MGD fails to download or reset. mgd_term windows fail to mount file system.
- The Carriage Fails to Advance to Latch to the Table and to Retract When it is Raised, Undocked or Lowered.
- Unable to Load Applications Software during restoreinfo Due to Corrupt saveinfo Disk.
- CWA ID100315. Auto and Manual Prescan Fails.
- Ghosted or Smeared Head Images Using Head Coil due to Head TR Assembly.
- A Copper Coolant Line or Pipe Internal to the MCS Cabinet Chiller Burst due to Water Remaining in Unit Freezing During Shipment.