GE Healthcare · MRExcite22692512269015
Software bug that is currently being investigated.
Symptom
The equipment room environmental sensor stopped collecting data following the 14.0 M5a SW FMI 60753. Both environmental sensors had been working prior to the sw upgrade.
Error message
2269251 The Equipment Room Temperature 1-Wire Device was not discovered on the 1-Wire Network Jan 27 Tue 06:48:31 2269251 1wireIfSerial 1wireUtil.c 1140 Exception Class: Notification Function: Peripherals: Communications The Equipment Room Temperature 1-Wire Device was not discovered on the 1-Wire Network.-> Extended Log: 2269251 ERROR: EM_ENVMON_1WIRE_EQUIP_NODE The Equipment Room Temperature 1-Wire Device was not discovered on the 1-Wire Network. Extended Error: This can happen if the device was not powered or properly connected when the system was booted. Verify the 1-Wire Device is properly connected. See system block diagrams if necessary. TPS Reset may be necessary to clear this error. 2269015 The information in the 1-Wire Device, Equipment Room Sensor, could not be successfully read. Verify the connections on the 1-Wire Network Jan 27 Tue 06:48:39 2269015 1wireIfSerial 1wireUtil.c 529 Exception Class: Notification Function: Software/Firmware/Config: N/A The information in the 1-Wire Device, Equipment Room Sensor, could not be successfully read. Verify the connections on the 1-Wire Network.-> Extended Log: 2269015 ERROR: EM_ENVMON_1WIRE_DEVICE_READ_ERR The information in the
Applies to
- Models
- Excite
- Product
- Excite HDx Includes "Excite HDi" system. 14.0 Software only.
- Software
- 14.0 M5 Excite HDx HDi Platforms only.
- Area
- MR
Full cause & step-by-step resolution
The complete cause analysis, the step-by-step remedy for this GE Healthcare MR issue are available to FSELIB subscribers — re-authored, web-first, searchable.
7-day free trial · no credit card required
Related GE Healthcare MR issues
- TPS/MGD fails to download or reset. mgd_term windows fail to mount file system.
- The Carriage Fails to Advance to Latch to the Table and to Retract When it is Raised, Undocked or Lowered.
- Unable to Load Applications Software during restoreinfo Due to Corrupt saveinfo Disk.
- CWA ID100315. Auto and Manual Prescan Fails.
- Ghosted or Smeared Head Images Using Head Coil due to Head TR Assembly.
- A Copper Coolant Line or Pipe Internal to the MCS Cabinet Chiller Burst due to Water Remaining in Unit Freezing During Shipment.