Signal Loss or Shading in one area of MRID 8ch Neurovascular coil 2317115-2 due to defective coil.

Symptom

Weak Grainy Images using the Eight (8) Channel NV Array Coil See Long Description and Attachment DWEPI with Asset images very grainy using 8NVHEAD NECK_A coil. Other coils okay and other PSDs okay. Error shown below was logged for this coil "Low Signal Detected". See Attachment for an example image. Wed Dec 10 16:11:08 2008 Error: 2224725 Low signal detected. Please check the signal level for receiver 0 (or receiver 1 while using Manual Prescan).

Error message

2224725 Low signal detected. Please check the signal level for receiver X (or receiver X while using Manual Prescan) See Long Description Error: 2224725 Host: hopgemr1a Process: NSP File: APS:aps_frame_range_check Line: 10 Low signal detected. Please check the signal level for receiver X (or receiver X while using Manual Prescan).

Applies to

Models
Excite
Product
Excite HD For 12.0 1.5T Systems only.
Software
12.0 Excite HD and 3T Excite HD products only
Component
SecondOpinion 1.5T 8ch NV Array 2317115-2 See comments in resolution concept for second opinion requirements before ordering this part
Area
MR

Full cause & step-by-step resolution