Fails TPS Reset when booting up to Applications or when Resetting TPS See Long Description When the MR system…

Power Surge or Drop.

Symptom

Fails TPS Reset when booting up to Applications or when Resetting TPS See Long Description When the MR system is booting up to applications it will fail TPS reset but Applications will successfully boot. With Applications up and "Reset TPS" has been selected, TPS fails to download successfully.

Error message

2237273 CERD BOOT ERROR CERD boot software on TPS timed out while waiting for a response from the CAP. Host: lx-bay14b Proc: NSP Error: 2237273 File: TPS:cx_dpr_utils.c Line: 196 CERD BOOT ERROR CERD boot software on TPS timed out while waiting for a response from the CAP. IT MAY BE NECESSARY TO CYCLE POWER ON THE ISE CHASSIS TO CLEAR THIS CONDITION. timeout duration = 100

2237297 CERD BOOT ERROR The CERD did not respond to a boot command in the allotted time. Host: lx-bay14b Proc: NSP Error: 2237297 File: TPS:cx_dpr_utils.c Line: 113 CERD BOOT ERROR The CERD did not respond to a boot command in the allotted time. CERD command : 32 response time : 100

2237285 CERD BOOT FAILURE Unable to load software to the CERD board. 2237285 Host: TSMR Process: NSP File: TPS:cx_init.c Line: 117 CERD BOOT FAILURE Unable to load software to the CERD board.

Applies to

Models
Signa, Ovation, Excite
Product
Signa LX Signa LX Product can be used with 8.x and 9.x systems.
Software
9.1 For Signa LX product only
Area
MR

Full cause & step-by-step resolution

The complete cause analysis, the step-by-step remedy for this GE Healthcare MR issue are available to FSELIB subscribers — re-authored, web-first, searchable.