Scan Pauses with Gradient Current Distortion and Gradient Data Overrange Errors, and Images Show a Partially Collapsed G…

Scan Pauses with Gradient Current Distortion and Gradient Data Overrange Errors, and Images Show a Partially Collapsed Gradient Field (White Line in Images) due to STIF Board.

Symptom

Scan Pauses with Gradient Current Distortion and Gradient Data Overrange Errors, and Images Show a Partially Collapsed Gradient Field (White Line in Images) See Long Description This is an HDe system running HDe 14.0 software. Scans are pausing with the errors shown below logged. When the customer resumes the scan, it may pause again with the same errors, the customer then resumes the scan again. When the scan is finished the images are reconstructed with a vertical line in these (axial in this example) images, which is indicative of a partially collapsed gradient field. See errors below from the site's message log, and the artifact that occurred (as a result of these exact errors) in the jmg attachment. You should be aware that the inclusion of the right arrow (-->) at the end of the message indicates that an extended error exists for that message with additional troubleshooting information. These extended errors can be seen by viewing the gesys* error log in Common Service Desktop. I've copied these extended errors into an attachment which is included here under the name "Extended_Errors". Tue Feb 10 10:29:29 2009 2244570 JSMR1 NSP SCP:Grad 1109 Exception Class: Unknown Severity: Unknown Function: Gradient: N/A PSD: FSE-XL Coil: 8CTL56 Scan: 460/8 The Z Axis gradient amplifier failed to go to Ready when commanded.-> Tue Feb 10 10:29:30 2009 2244851 JSMR1 NSP SCP:Grad 2870 Exception Class: Unknown Severity: Unknown Function: Gradient: N/A PSD: FSE-XL Coil: 8CTL56 Scan: 460/8 The GP is reporting a Gradient Amplifier Z Axis Current Distortion Fault.-> Tue Feb 10 10:30:54 2009 2250082 JSMR1 NSP SCP:Grad 3255 Exception Class: Unknown Severity: Unknown Function: Gradient: N/A PSD: FSE-XL Coil: 8CTL56 Scan: 460/8 Z Axis Gradient Data Overrange Occurred.->

Error message

2250082 Z Axis Gradient Data Overrange Occurred. 2250082 MD-DMR-MRGE2 NSP SCP:Grad 3253 Exception Class: Unknown Severity: Unknown Function: Gradient: N/A PSD: Fiesta Coil: 8Ch Body Upper Scan: 4/15 Z Axis Gradient Data Overrange Occurred.->

2244570 The gradient amplifier failed to go to Ready when commanded. 2244570 TOiMR2 NSP SCP:Grad 1106 Exception Class: Unknown Severity: Unknown Function: Gradient: N/A PSD: Spin Echo Coil: US SHOULDER PA Scan: 379/4 The X Axis gradient amplifier failed to go to Ready when commanded.-> 2244570 TOiMR2 NSP SCP:Grad 1106 Exception Class: Unknown Severity: Unknown Function: Gradient: N/A PSD: Spin Echo Coil: US SHOULDER PA Scan: 379/4 The X Axis gradient amplifier failed to go to Ready when commanded.

2244851 The GP is reporting a Gradient Amplifier %c Axis Current Distortion Fault. the same error as 2244577 on 11X, G3, openMR4 and 2244578 on 12X, E2, 14M4 (HDx, HDe), and openMR5. See Long Description 2244851 error code on Post 14M5 and 12.0 M5, is the same error as 2244577 on 11X, G3, openMR4and 2244578 on 12X, E2, 14M4 (HDx, HDe), and openMR5.2244851 The GP is reporting a Gradient Amplifier %c Axis Current Distortion Fault Example error message

Applies to

Models
Excite
Product
Excite HDe. For Hino HDe 14.0 and HDe 15.0 Software Systems only
Software
HDe 14.0 For Hino Excite HDe 14.0 systems only.
Component
2280952-3 STIF Board See Long Description SRF TRF I/O or STIF 2280952 crosses up in FEMC to 2280952-3. Part Description is "STIF Assembly".