Scans Stop Intermittently with Spurious Dynamic Disable and T/R Driver Board Error due to System Support Module SSM.
Symptom
Scans Stop Intermittently with Spurious Dynamic Disable and T/R Driver Board Error. See Long Description This is a 1.5T mobile Goldseal system running 11.1 software that had this problem since the mobile arrived in the FE's area. The customer when scanning is getting intermittent spurious Dynamic disable and T/R driver board error. It happens with all coils and can't be pinned down to a specific PSD. See below for the error that is occurring. 2249157 Dynamic disable and T/R driver board error. A spurious dynamic disable fault occurred. It cleared itself before the specific problem could be read. Possible causes: noise on the dynamic disable board or a bad board.
Error message
2249157 Dynamic Disable And T/R Driver Board Error Wed Aug 4 10:56:19 2004 Error: 2249157 Host: SIG1_OC0 Process: NSP File: SCP:../TrDriverMdsHWO.cpp Line: 1360 Dynamic disable and T/R driver board error. A spurious dynamic disable fault occurred. It cleared itself before the specific problem could be read. Possible causes: Noise on the dynamic disable board or a bad board.
Applies to
- Models
- Excite, Signa
- Product
- Excite Includes Excite 1 and Excite 2, 10.0, 11.0 and 11.1 Software.
- Software
- 11.1 For product 1.5T Excite only. Note, 3T Excite goes with software G3, not 11.0 or 11.1.
- Component
- SecondOpinion System Support Module 2160200-43 See comments in resolution concept for second opinion requirements before ordering this part. FE replaced SSM which is now a FRU Part #: 2160200-43 Part Type: GP REPAIRABLES Part Weight: 48.50 LBR Repairable: Yes Returnable: Yes Hazardous Material: N Description: ERBTEC #550020 SYSTEM SUPPORT MODULE FRU: Yes Status: Orderable ICV/Transfer Price: 5684.00 USD /Each List Price: 36215.22 USD /EACH See comments in resolution concept for second opinion requirements before ordering this part. FE replaced SSM which is now a FRU Part #: 2160200-43 Part Type: GP REPAIRABLES Part Weight: 48.50 LBR Repairable: Yes Returnable: Yes Hazardous Material: N Description: ERBTEC #550020 SYSTEM SUPPORT MODULE FRU: Yes Status: Orderable ICV/Transfer Price: 5684.00 USD /Each List Price: 36215.22 USD /EACH
- Area
- MR
Full cause & step-by-step resolution
The complete cause analysis, the step-by-step remedy for this GE Healthcare MR issue are available to FSELIB subscribers — re-authored, web-first, searchable.
7-day free trial · no credit card required
Related GE Healthcare MR issues
- Running grafidy3, and find there are spikes on the waveforms in the manual prescan display, both windows.
- TPS/MGD fails to download or reset. mgd_term windows fail to mount file system.
- Fails to Scan with TORSO Coil.
- The Carriage Fails to Advance to Latch to the Table and to Retract When it is Raised, Undocked or Lowered.
- AE title has changed on destination host.
- Unable to Load Applications Software during restoreinfo Due to Corrupt saveinfo Disk.