CWA ID 126542 Dot Artifact Sometimes Seen In Exact Center of Images Acquired When Using GE_HD 8CH Shoulder Coil.

Symptom

Dot Artifact Sometimes Seen In Exact Center of Images Acquired When Using GE_HD 8CH Shoulder Coil See Long Description Dot artifact sometimes seen in exact center of the images acquired when using GE_HD 8CH Shoulder coil. There are no errors logged. See attached image for image showing shoulder coil dot artifact. CIM/CSO # 1-1658524504, iTrak # 13238441 This artifact is the same as referenced in SMILe dated 12/01/2005, see below, but is seen on another coil. SMILe 12/01/2005 Subject: Dot Artifact Sometimes Seen In Middle of Images Acquired When Using 1.5T HD Knee Array Coil.

Applies to

Models
Excite
Product
Excite HDx Includes "Excite HDi" system. 14.0 Software only.
Software
14.0 M5 Excite HDx HDi Platforms only.
Component
2274188-3 RRF MULTIPLEXER WITH LOW PASS FILTER
Area
MR