BrainwaveRT fMRI and RTIA (iDrive) fail both prescan and scan with errors logged.
Symptom
BrainwaveRT fMRI and RTIA (iDrive) Fail both Prescan and Scan See Long Description The prescan and scan for BrainwaveRT fMRI and RTIA (iDrive) always abort and log the following errors in the message log. All diagnostics pass and AGP, APS, AP, DRF boards all swapped with no effect on the problem. Wed Jun 2 20:59:23 2004 2218537 ex-mr scn MsaScanMgr.m 2641 Scan stopped, Got Recon-Scan Stop packet with error message key - 2227854. Wed Jun 2 20:59:23 2004 2227834 ex-mr NSP APS:da_dacq_task 21 A data acquisition error occurred. Run datapath diagnostics. 00000068H Wed Jun 2 20:59:23 2004 2248014 ex-mr NSP APS:DRF 262 Allocate DPR memory error. Possible reason: Too many printf or log msg caused APS missed EOS interrupt. Software error. Please change prescription or perform a TPS Reset. DRF ID: 4 FE ID: 0 arg0: 1 arg1: 0
Error message
2227834 A Data Acquisition Error Occurred 2227834 1 ex-mr NSP APS:da_dacq_task 21 A data acquisition error occurred. Run datapath diagnostics. AP board level diagnostics, and Image Transfer (IT) board level diagnostics. Multiple iterations may be required before the error happens. 00000138H EN 314 ER 395 2227834 1 ex-mr NSP APS:da_dacq_task 21 A data acquisition error occurred. Run datapath diagnostics. 00000068H 2248014 Allocate DPR memory error 2248014 - Allocate DPR memory error. Possible reason : too many printf or log msg caused APS missed EOS interrupt. Software error. Please change prescription or perform a TPS reset. DRF ID : 4 FE ID : 0 arg0 : 1 arg1 : 0 2218537 Scan stopped, Got Recon-Scan Stop packet With Error Message Key Fri Apr 16 10:11:34 2004 Error: 2218537 Host: LX1 Process: scn File: MsaScanMgr.m Line: 2641 Scan stopped, Got Recon-Scan Stop packet with error message key - 2248
Applies to
- Models
- Excite
- Product
- Excite Includes Excite 1 and Excite 2, 10.0, 11.0 and 11.1 Software.
- Software
- 11.0 Only for 1.5T Excite product. 12.0 is Excite HD and 14.0 is Excite HDx. Note, 3T Excite goes with software G3, not 11.0 or 11.1.
- Area
- MR
Full cause & step-by-step resolution
The complete cause analysis, the step-by-step remedy for this GE Healthcare MR issue are available to FSELIB subscribers — re-authored, web-first, searchable.
7-day free trial · no credit card required
Related GE Healthcare MR issues
- TPS/MGD fails to download or reset. mgd_term windows fail to mount file system.
- The Carriage Fails to Advance to Latch to the Table and to Retract When it is Raised, Undocked or Lowered.
- Unable to Load Applications Software during restoreinfo Due to Corrupt saveinfo Disk.
- CWA ID100315. Auto and Manual Prescan Fails.
- Ghosted or Smeared Head Images Using Head Coil due to Head TR Assembly.
- A Copper Coolant Line or Pipe Internal to the MCS Cabinet Chiller Burst due to Water Remaining in Unit Freezing During Shipment.