BrainwaveRT fMRI and RTIA (iDrive) fail both prescan and scan with errors logged.

Symptom

BrainwaveRT fMRI and RTIA (iDrive) Fail both Prescan and Scan See Long Description The prescan and scan for BrainwaveRT fMRI and RTIA (iDrive) always abort and log the following errors in the message log. All diagnostics pass and AGP, APS, AP, DRF boards all swapped with no effect on the problem. Wed Jun 2 20:59:23 2004 2218537 ex-mr scn MsaScanMgr.m 2641 Scan stopped, Got Recon-Scan Stop packet with error message key - 2227854. Wed Jun 2 20:59:23 2004 2227834 ex-mr NSP APS:da_dacq_task 21 A data acquisition error occurred. Run datapath diagnostics. 00000068H Wed Jun 2 20:59:23 2004 2248014 ex-mr NSP APS:DRF 262 Allocate DPR memory error. Possible reason: Too many printf or log msg caused APS missed EOS interrupt. Software error. Please change prescription or perform a TPS Reset. DRF ID: 4 FE ID: 0 arg0: 1 arg1: 0

Error message

2227834 A Data Acquisition Error Occurred 2227834 1 ex-mr NSP APS:da_dacq_task 21 A data acquisition error occurred. Run datapath diagnostics. AP board level diagnostics, and Image Transfer (IT) board level diagnostics. Multiple iterations may be required before the error happens. 00000138H EN 314 ER 395 2227834 1 ex-mr NSP APS:da_dacq_task 21 A data acquisition error occurred. Run datapath diagnostics. 00000068H

2248014 Allocate DPR memory error 2248014 - Allocate DPR memory error. Possible reason : too many printf or log msg caused APS missed EOS interrupt. Software error. Please change prescription or perform a TPS reset. DRF ID : 4 FE ID : 0 arg0 : 1 arg1 : 0

2218537 Scan stopped, Got Recon-Scan Stop packet With Error Message Key Fri Apr 16 10:11:34 2004 Error: 2218537 Host: LX1 Process: scn File: MsaScanMgr.m Line: 2641 Scan stopped, Got Recon-Scan Stop packet with error message key - 2248

Applies to

Models
Excite
Product
Excite Includes Excite 1 and Excite 2, 10.0, 11.0 and 11.1 Software.
Software
11.0 Only for 1.5T Excite product. 12.0 is Excite HD and 14.0 is Excite HDx. Note, 3T Excite goes with software G3, not 11.0 or 11.1.
Area
MR

Full cause & step-by-step resolution