MGD Power Supply PS2 Replaced But New Power Supply Has Same Problem due to the Actual Failing MGD Power Supply Was The O…

MGD Power Supply PS2 Replaced But New Power Supply Has Same Problem due to the Actual Failing MGD Power Supply Was The One With No Fault LED Lit.

Symptom

MGD Power Supply PS2 Replaced But New Power Supply Has Same Problem See Long Description Every MGD chassis has two matching power supplies in its front plane: PS1 and PS2. These supplies have Fault LEDs that will come on if there is a problem within the supply. At this site the FE replaced the PS2 MGD Power Supply because one of its Fault LEDs was lit. PS1 did not have any Fault LED lit. But when he put the new PS2 power supply in, the same Fault LED was immediately lit on the new PS2 just like on the original one. The FE called the OnLine Center to ask of there was a general problem with these supplies, and might he have gotten a bad one from stock?

Applies to

Models
Excite
Product
3T Excite HDx 14.0 Software. The software that this solution is tied to is indicated by the Software concept below.
Software
14.0 M5 Excite HDx HDi Platforms only.
Area
MR

Full cause & step-by-step resolution