GE Healthcare · MRExcite2254270117472254236
2254270 3T RF Amplifier IPA Tube Arc (fault 67) due to MKS 35kW RF Amp.
Symptom
Unable to Scan with 3T MKS RF Amplifier will Not go Ready MKS amp powers up but will not come ready.
Error message
2254270 3T RF Amplifier IPA Tube Arc (fault 67). Inter-electrode arc within tube was detected. Error: 2254270 Host: gh3t-ow1 Process: NSP File: SCP:../rf nb contro Line: 1491 Exception Class: Unknown Severity: Unknown Function: RF: RF Transmit PSD: Gradient Echo Coil: HDCTL 456 Scan: 11747/1 3T RF Amplifier IPA Tube Arc (fault 67). Inter-electrode arc within tube was detected. Possible Causes: spurious arc, defective IPA tube. If the fault persists, replace the RF amplifier 2254236 RF Amplifier PA High Voltage Supply Low (fault 15). 2254236 Host: EHRMR3T Process: NSP File: SCP:../RfNbMdsHWO.cpp Line: 2505 Exception Class: Unknown Severity: Unknown Function: RF: RF Transmit PSD: Gradient Echo Coil: 8US TORSOPA Scan: 4283/1 3T RF Amplifier PA High Voltage Supply Low (fault 15). Possible Causes: defective PA Tube, defective PA HV supply. If fault persists, replace the RF amplifier.-
Applies to
- Models
- Excite
- Product
- 3T Excite HDx 14.0 Software. The software that this solution is tied to is indicated by the Software concept below.
- Software
- 14.0 M5 Excite HDx HDi Platforms only.
- Component
- SecondOpinion MKS 3T 35kW RF Amp 5104896 See comments in resolution concept for second opinion requirements before ordering this part.
- Area
- MR
Full cause & step-by-step resolution
The complete cause analysis, the step-by-step remedy for this GE Healthcare MR issue are available to FSELIB subscribers — re-authored, web-first, searchable.
7-day free trial · no credit card required
Related GE Healthcare MR issues
- TPS/MGD fails to download or reset. mgd_term windows fail to mount file system.
- The Carriage Fails to Advance to Latch to the Table and to Retract When it is Raised, Undocked or Lowered.
- Unable to Load Applications Software during restoreinfo Due to Corrupt saveinfo Disk.
- CWA ID100315. Auto and Manual Prescan Fails.
- Ghosted or Smeared Head Images Using Head Coil due to Head TR Assembly.
- A Copper Coolant Line or Pipe Internal to the MCS Cabinet Chiller Burst due to Water Remaining in Unit Freezing During Shipment.