GE Healthcare · MRExcite2252302

Unable to Scan any Multicoil due to Error 2252302 Multicoil Bias Fault Transmit Fault Always on Channel 1 due to Pin J1…

Unable to Scan any Multicoil due to Error 2252302 Multicoil Bias Fault Transmit Fault Always on Channel 1 due to Pin J1 Retracted or Pushed Inside Port A of LPCA.

Symptom

Unable to Scan any Multi coil due to Error 2252302 Multicoil Bias Fault - Transmit Fault Always on Channel 1 See Long Description Unable to scan using any of the site's multicoils due to the Multicoil Bias Fault - Transmit Fault shown below. But any non-multicoil such as Head and Body work fine. 2252302 WCLAGEMR2MR NSP SCP:DCB 787 Exception Class: Unknown Severity: Unknown Function: RF: RF Bias PSD: FGRE-XL Coil: HDNV Head Scan: 50653/1 Multicoil Bias Fault - Transmit Fault. Try re-seating all coil connectors. MCD1 Channel 1 of 16, counting from 1 <-- only the 1st channel failure is reported Value: -2 mA, Limit: 0 mA Mask of failed channels: 0x1 <-- in this case the mask of 0x1 tells us that only channel 1 is failing MCD1 1-8 : A or C MCD1 9-16 : B MCD2 1-8 : B or Legacy MC MCD2 9-16 : C->

Error message

2252302 Multicoil Bias Fault Detected - Transmit Fault 2252302 IC_OPEN_SG NSP SCP:DCB 793 Multicoil Bias Fault Detected - Transmit Fault. Try re-seating all coil connectors. MCD1 Channel 5 of 16, counting from 1 Value: -2 mA Limit: 0 mA Mask of failed channels: 0x10 Ch 1-8 : Legacy MC Conn Ch 17-24 : Conn A Ch 25-32 : Conn B->

Applies to

Models
Excite
Product
Excite HDe. For Hino HDe 14.0 and HDe 15.0 Software Systems only
Software
HDe 14.0 For Hino Excite HDe 14.0 systems only.
Area
MR

Full cause & step-by-step resolution

The complete cause analysis, the step-by-step remedy for this GE Healthcare MR issue are available to FSELIB subscribers — re-authored, web-first, searchable.