GE Healthcare · MRExcite808857866921564
ermes Errors Due to Corrupt Error Message Database.
Symptom
System is Logging ermes Errors With No Apparent Effect on System. The "ermes" errors of the type shown below are being logged, filling up the message log. The error messages don't explain what's wrong. There doesn't seem to be any effect on the system otherwise. Mon Feb 11 13:29:11 2008 0 4 lx-mr2 scn msgFile.c 315 The message key -4 was not found. Mon Feb 11 13:29:19 2008 -4 4 lx-mr2 scn PsdError.m 377 Ermes error, key: -4, error: 4009.
Error message
0 The Ermes message file is inconsistent, it must be rebuilt. Wed Mar 14 06:00:44 2001 Error: 0 Host: MR01_OW0 Process: hostMonitor File: msgFile.c Line: 259 The Ermes message file is inconsistent, it must be rebuilt. Detail: 808857866 921564 0 0 0 0
Applies to
- Models
- Excite
- Product
- Excite Includes Excite 1 and Excite 2, 10.0, 11.0 and 11.1 Software.
- Software
- 11.0 Only for 1.5T Excite product. 12.0 is Excite HD and 14.0 is Excite HDx. Note, 3T Excite goes with software G3, not 11.0 or 11.1.
- Area
- MR
Full cause & step-by-step resolution
The complete cause analysis, the step-by-step remedy for this GE Healthcare MR issue are available to FSELIB subscribers — re-authored, web-first, searchable.
7-day free trial · no credit card required
Related GE Healthcare MR issues
- TPS/MGD fails to download or reset. mgd_term windows fail to mount file system.
- The Carriage Fails to Advance to Latch to the Table and to Retract When it is Raised, Undocked or Lowered.
- Unable to Load Applications Software during restoreinfo Due to Corrupt saveinfo Disk.
- CWA ID100315. Auto and Manual Prescan Fails.
- Ghosted or Smeared Head Images Using Head Coil due to Head TR Assembly.
- A Copper Coolant Line or Pipe Internal to the MCS Cabinet Chiller Burst due to Water Remaining in Unit Freezing During Shipment.