GE Healthcare · MRExcite2249147
High Voltage Supply drops to around 500 volts in SSM when trying to scan with surface coils.Unable to scan with all surf…
High Voltage Supply drops to around 500 volts in SSM when trying to scan with surface coils.Unable to scan with all surface coils.
Symptom
Surface Coil Scans Fail with SSM High Voltage Error With any suface coil and a high duty cycle scan.
Error message
2249147 High Voltage Power Supply Low. Also may be a shorted load on dynamic disable or direct drive circuits. Tue Mar 30 13:50:47 2004 Error: 2249147 Host: CCRC Process: NSP File: SCP:../TrDriverMdsHWO.cpp Line: 1255 SSM fault.^A High voltage power supply low. Also may be a shorted load on dynamic disable or direct drive circuits.
Applies to
- Models
- Excite
- Product
- Excite Includes Excite 1 and Excite 2, 10.0, 11.0 and 11.1 Software.
- Software
- 11.0 Only for 1.5T Excite product. 12.0 is Excite HD and 14.0 is Excite HDx. Note, 3T Excite goes with software G3, not 11.0 or 11.1.
- Component
- 2160200-10 SSM HV Power Supply
- Area
- MR
Full cause & step-by-step resolution
The complete cause analysis, the step-by-step remedy for this GE Healthcare MR issue are available to FSELIB subscribers — re-authored, web-first, searchable.
7-day free trial · no credit card required
Related GE Healthcare MR issues
- TPS/MGD fails to download or reset. mgd_term windows fail to mount file system.
- The Carriage Fails to Advance to Latch to the Table and to Retract When it is Raised, Undocked or Lowered.
- Unable to Load Applications Software during restoreinfo Due to Corrupt saveinfo Disk.
- CWA ID100315. Auto and Manual Prescan Fails.
- Ghosted or Smeared Head Images Using Head Coil due to Head TR Assembly.
- A Copper Coolant Line or Pipe Internal to the MCS Cabinet Chiller Burst due to Water Remaining in Unit Freezing During Shipment.