Invivo 8 Ch Wrist Array MCQA SNR Test Failure due to Known Software Coding Issue in 14.0 M5 Software.

Symptom

Invivo 8 Ch Wrist Array Fails MCQA Tool SNR Test. Specific Channel Failures Not Repeatable. See Long Description When the MCQA test is run on this coil multiple times, channels that were failing will pass and other channels that were passing will fail. No errors posted to error log. Replacing coil has no effect. MCQA fails intermittently. itrak # 13103152 spr # MRIhc23214

Applies to

Models
Excite
Product
Excite HDx Includes "Excite HDi" system. 14.0 Software only.
Software
14.0 M5 Excite HDx HDi Platforms only.
Area
MR

Full cause & step-by-step resolution