GE Healthcare · X-ray VascularInnova42000674200035400015
125937 Innova 2000 M4b known service limitations
Symptom
Innova 2000 M4b known service limitations M4b has several issues that effect service including bogus errors in the error log, bogus boot up errors. Diagnostic and calibration limitations. These errors may display on bootup and do not effect system operation. 4AC5 4ACA 4F28 4F2C 4F30 4F34 4F3C These are operating errors that do not effect operation: 4AC5 4AC3 4105 20005 810007 1610007 These calibrations are affected P011 X220
Error message
error Code = 4200067, error Text = User request failed followed by error Code = 4200035, error Text = FOV Notification timer has expired followed by error Code = 400015, error Text = Time out while in , waiting for, System impact: Exam failure. Displayed message : "Selection failed, retry protocol selection", "Xray not available - Reset system; if problem persists, call Service".
Applies to
- Models
- Innova
- Product
- Innova 2000 Innova 2000 systems Innova 2000 systems
- Software
- $, Innova 2000 H2VCP M4 FMI, DL release M4_12.17_REV1 M4_12.17_REV1 DL release 12.32.2 Ghost DL-1 "F" : NA Ghost DL-1 "G" : NA Ghost DL-2 : NA Ghost DL-4 : 19.9 DL : 12.32.2 DL Atlas Core : 3.1Q RTAC Atlas Core : 3.1O-1 IDC : 1.15 Calypso : 13.6 Titan : H0.58 Pos. Motorola : NA Pos. Roadrunner : 5.2K-2 Calypso Uptime : 12.17_Rev1
- Area
- X-ray Vascular
Full cause & step-by-step resolution
The complete cause analysis, the step-by-step remedy for this GE Healthcare X-ray Vascular issue are available to FSELIB subscribers — re-authored, web-first, searchable.
7-day free trial · no credit card required
Related GE Healthcare X-ray Vascular issues
- The Worklist does not refresh on the system when performing a query.
- Innova System boot failure
- INX1X; Message Limit switch engaged; LP4 Lateral Detector lift will not move after going to max SID using the Emergency Backout button on the lift cover because the lift hit maximum (Max) SID limit sw…
- Firewall is broadcasting DHCP addresses
- Cooling Failure; Shut System down and call Service is displayed on the Innova Image monitor
- No gantry motion is reported by the customer.