Pular para o conteúdo principal
GE Healthcare · X-ray R&FAdvantx2108822

ADXE ECM module (2111390-2) known diagnostic errors

Symptom

Need the known diagnostic errors with the Advantx-E ECM module part number 2111390-2

Error message

8221 User 001F Hardware codes 3300, 3400, 3500, 3600 8221  Timer Operation Test Failure Hdwr status 3300 User status 001f 8221  Timer Operation Test Failure Hdwr status 3400 User status 001f 8221  Timer Operation Test Failure Hdwr status 3500 User status 001f 8221  Timer Operation Test Failure Hdwr status 3600 User status 001f These errors are generated from the use of the CTS9513 timer at location U10 and U31 on the 2108822-2 board inside the ecm enclosure.

Applies to

Models
Advantx
Product
DRS Advantx-E
Area
X-ray R&F

Full cause & step-by-step resolution

The complete cause analysis, the step-by-step remedy for this GE Healthcare X-ray R&F issue are available to FSELIB subscribers — re-authored, web-first, searchable.

7-day free trial · no credit card required