GE Healthcare · X-ray VascularInnova5000011
Boot failure with error 5000011, System Boot Failed waiting for boot to complete
Symptom
Boot failure with error 5000011, System Boot Failed waiting for boot to complete The system will not boot with the hospital network cable connected If the hospital network cable is connected the system fail to boot with XRSSubSystemMGRDL process time out. Error 5000011. If you disconnect the network cable the system boots fine. You can then re-connect the network cable and the system continues to work properly and is able to transfer images to the PACS system.
Error message
5000011 "System Boot Failed: Timed Out waiting for boot to complete, review the debug text to determine which actors (module) is offline. The system has to be Reset
Applies to
- Models
- Innova
- Product
- Innova X100 Monoplane Innova X100 Monoplane from product name Innova 2100, 3100 and 4100 Innova X100 Monoplane Innova 2100, 3100 and 4100 Innova 2000 Innova 2000 systems Innova 2000 systems Innova X1 X1 Biplane Innova X1 X1 Bi-plane Innova 2121 or 3131 Bi-plane System Innova X1 X1 Bi-plane Innova 2121 or 3131 Bi-plane System
- Area
- X-ray Vascular
Full cause & step-by-step resolution
The complete cause analysis, the step-by-step remedy for this GE Healthcare X-ray Vascular issue are available to FSELIB subscribers — re-authored, web-first, searchable.
7-day free trial · no credit card required
Related GE Healthcare X-ray Vascular issues
- The Worklist does not refresh on the system when performing a query.
- Innova System boot failure
- INX1X; Message Limit switch engaged; LP4 Lateral Detector lift will not move after going to max SID using the Emergency Backout button on the lift cover because the lift hit maximum (Max) SID limit sw…
- Firewall is broadcasting DHCP addresses
- Cooling Failure; Shut System down and call Service is displayed on the Innova Image monitor