Pular para o conteúdo principal
GE Healthcare · MRExcite2250083225007951000

Scan Pauses with Gradient Errors Due to STIF board.

Symptom

Scan Pauses with Gradient Errors and Continues after start Scan Pressed then Images Show Bright Line Artifact System scans and pauses with Gradient errors. Images had a definite gradient collapsed.

Error message

2250083 Y Axis Gradient Data Overrange Occurred. Error: 2250083 Host: CHCMR2 Process: NSP File: SCP:Grad Line: 3255 Exception Class: Unknown Severity: Unknown Function: Gradient: N/A PSD: FSE-XL Coil: HD TRknee PA Scan: 2458/7 Y Axis Gradient Data Overrange Occurred.-

2250079 % Axis Gradient Data Overrange Occurred see long description Sun Feb 3 23:48:08 2008 Error: 2250079 Host: clmr2 Process: NSP File: SCP:Grad Line: 3255 Exception Class: Unknown Severity: Unknown Function: Gradient: N/A PSD: fgrespt Coil: HEAD Scan: 51000/1 X Axis Gradient Data Overrange Occurred.-> Extended Error: 2250079 ERROR: EM_GP_X_SGA_OVRANGE_FAULT X Axis Gradient Data Overrange Occurred. Extended Error: This condition is noncritical and should not affect image quality, unless accompanied by additional error messages. An occurrence of this type, unless accompanied by additional error messages, is not due to hardware fault. If this error occurs frequently, please submit a PQR describing the conditions under which the error occurs.

Applies to

Models
Excite
Product
Excite HD For 12.0 1.5T Systems only.
Software
12.0 Excite HD and 3T Excite HD products only
Component
2280952-3 STIF Board See Long Description SRF TRF I/O or STIF 2280952 crosses up in FEMC to 2280952-3. Part Description is "STIF Assembly".
Area
MR

Full cause & step-by-step resolution

The complete cause analysis, the step-by-step remedy for this GE Healthcare MR issue are available to FSELIB subscribers — re-authored, web-first, searchable.