GE Healthcare · MRExcite2250083225007951000
Scan Pauses with Gradient Errors Due to STIF board.
Symptom
Scan Pauses with Gradient Errors and Continues after start Scan Pressed then Images Show Bright Line Artifact System scans and pauses with Gradient errors. Images had a definite gradient collapsed.
Error message
2250083 Y Axis Gradient Data Overrange Occurred. Error: 2250083 Host: CHCMR2 Process: NSP File: SCP:Grad Line: 3255 Exception Class: Unknown Severity: Unknown Function: Gradient: N/A PSD: FSE-XL Coil: HD TRknee PA Scan: 2458/7 Y Axis Gradient Data Overrange Occurred.- 2250079 % Axis Gradient Data Overrange Occurred see long description Sun Feb 3 23:48:08 2008 Error: 2250079 Host: clmr2 Process: NSP File: SCP:Grad Line: 3255 Exception Class: Unknown Severity: Unknown Function: Gradient: N/A PSD: fgrespt Coil: HEAD Scan: 51000/1 X Axis Gradient Data Overrange Occurred.-> Extended Error: 2250079 ERROR: EM_GP_X_SGA_OVRANGE_FAULT X Axis Gradient Data Overrange Occurred. Extended Error: This condition is noncritical and should not affect image quality, unless accompanied by additional error messages. An occurrence of this type, unless accompanied by additional error messages, is not due to hardware fault. If this error occurs frequently, please submit a PQR describing the conditions under which the error occurs.
Applies to
- Models
- Excite
- Product
- Excite HD For 12.0 1.5T Systems only.
- Software
- 12.0 Excite HD and 3T Excite HD products only
- Component
- 2280952-3 STIF Board See Long Description SRF TRF I/O or STIF 2280952 crosses up in FEMC to 2280952-3. Part Description is "STIF Assembly".
- Area
- MR
Full cause & step-by-step resolution
The complete cause analysis, the step-by-step remedy for this GE Healthcare MR issue are available to FSELIB subscribers — re-authored, web-first, searchable.
7-day free trial · no credit card required
Related GE Healthcare MR issues
- TPS/MGD fails to download or reset. mgd_term windows fail to mount file system.
- The Carriage Fails to Advance to Latch to the Table and to Retract When it is Raised, Undocked or Lowered.
- Unable to Load Applications Software during restoreinfo Due to Corrupt saveinfo Disk.
- CWA ID100315. Auto and Manual Prescan Fails.
- Ghosted or Smeared Head Images Using Head Coil due to Head TR Assembly.
- A Copper Coolant Line or Pipe Internal to the MCS Cabinet Chiller Burst due to Water Remaining in Unit Freezing During Shipment.