GE Healthcare · MRExcite224837122483702218537

Three 3-Plane Localizers Produce Only Three 3 Images Per Plane and Scans Prescribed From Them Abort With SCT Interface E…

Three 3-Plane Localizers Produce Only Three 3 Images Per Plane and Scans Prescribed From Them Abort With SCT Interface Errors due to Obsolete Infiniband Firmware Version.

Symptom

Three 3-Plane Localizers Produce Only Three 3 Images Per Plane and Scans Prescribed From Them Abort With SCT Interface Errors See Long Description FMI 60753 (14.0 M5A software upgrade) installed Thursday evening, January 15, 2009. Upon the customer beginning their scanning day on Friday, January 16 they noticed the following problem: The 3-Plane Loc's produced only one image per plane, and when they tried to prescribe off these images and scan the first axial scan, the scans would always abort with the SCT Interface Errors as shown below. iTrak # 13236909. Fri Jan 16 08:25:35 2009 2248370 mcoamr1 NSP SCP:SctPxProxy 833 Exception Class: Unknown Severity: Unknown Function: Software/Firmware/Config: N/A PSD: Gradient Echo Coil: QUADANKLE Scan: 5207/1 SCT Interface Error. No source was detected for a SCT command. Fri Jan 16 08:25:35 2009 2248371 mcoamr1 NSP SCP:SctPxProxy 854 Exception Class: Unknown Severity: Unknown Function: Software/Firmware/Config: N/A PSD: Gradient Echo Coil: QUADANKLE Scan: 5207/1 SCT Interface Error. No timestamp was detected for a SCT command. Fri Jan 16 08:25:44 2009 2218537 mcoamr1 scn MsaScanMgr.m 3038 Scan stopped, Scan Stopped due to error : SCT Interface Error. No source was detected for a SCT command..

Error message

2248371 SCT Interface Error. No timestamp was detected for a SCT command. Tue Mar 23 08:08:27 2004 Error: 2248371 Host: IRIS Process: NSP File: SCP:SctPxProxy Line: 538 SCT Interface Error. No timestamp was detected for a SCT command. Tue Mar 23 08:09:31 2004 Error: 2248370 Host: IRIS Process: NSP File: SCP:SctPxProxy Line: 653 SCT Interface Error. No source was detected for a SCT command.

2248370 SCT Interface Error. No source was detected for a SCT command. See Long Description 2248370 mcoamr1 NSP SCP:SctPxProxy 833 Exception Class: Unknown Severity: Unknown Function: Software/Firmware/Config: N/A PSD: Gradient Echo Coil: QUADANKLE Scan: 5207/1 SCT Interface Error. No source was detected for a SCT command

2218537 Scan stopped, Scan Stopped due to error SCT Interface Error. No source was detected for a SCT command. See Long Description 2218537 mcoamr1 scn MsaScanMgr.m 3038 Scan stopped, Scan Stopped due to error : SCT Interface Error. No source was detected for a SCT command..

Infiniband backlog too large…scan stopped See Long Description “ Infiniband backlog too large…scan stopped " Message is seen in the APS window.

Applies to

Models
Excite
Product
Excite HDx Includes "Excite HDi" system. 14.0 Software only.
Software
14.0 M5 Excite HDx HDi Platforms only.
Area
MR

Full cause & step-by-step resolution

The complete cause analysis, the step-by-step remedy and the attached service documents for this GE Healthcare MR issue are available to FSELIB subscribers — re-authored, web-first, searchable.