GE Healthcare · MRExcite2248673
The ACGD 3-Phase Fan Assembly is shorted.
Symptom
ACGD Fan Assembly trips ACGD 208 VAC breaker in PDU.
Error message
2248673 MDS link failure. The MDS fiber optic link appears to be broken between the Gradient and the RF cabinets. 2248673 Host: lxmr Process: NSP File: SCP:MdsLinkHWO Line: 1713 MDS link failure. The MDS fiber optic link appears to be broken between the Gradient and the RF cabinets. Check gradient cabinet power, fiber optic connections and signal level of fiber optic link. Run MDS fiber optic link diagnostics.
Applies to
- Models
- Excite
- Product
- Excite Includes Excite 1 and Excite 2, 10.0, 11.0 and 11.1 Software.
- Software
- 11.0 Only for 1.5T Excite product. 12.0 is Excite HD and 14.0 is Excite HDx. Note, 3T Excite goes with software G3, not 11.0 or 11.1.
- Component
- 2299215 ACGD CABINET FAN DIRECTION 2138247 pages 15-2, 15-4, 15-8
- Area
- MR
Full cause & step-by-step resolution
The complete cause analysis, the step-by-step remedy for this GE Healthcare MR issue are available to FSELIB subscribers — re-authored, web-first, searchable.
7-day free trial · no credit card required
Related GE Healthcare MR issues
- TPS/MGD fails to download or reset. mgd_term windows fail to mount file system.
- The Carriage Fails to Advance to Latch to the Table and to Retract When it is Raised, Undocked or Lowered.
- Unable to Load Applications Software during restoreinfo Due to Corrupt saveinfo Disk.
- CWA ID100315. Auto and Manual Prescan Fails.
- Ghosted or Smeared Head Images Using Head Coil due to Head TR Assembly.
- A Copper Coolant Line or Pipe Internal to the MCS Cabinet Chiller Burst due to Water Remaining in Unit Freezing During Shipment.