Paused scans and image artifacts due to Power sagging from the normal 480 VAC down to as low as 0 VAC during the pulse s…

Paused scans and image artifacts due to Power sagging from the normal 480 VAC down to as low as 0 VAC during the pulse sequence.

Symptom

MRIQ Scan stops Multiple Times, They Restart the Scan and it Completes but Images Have Severe Zipper Artifacts. http://3.87.118.28/optec3/common/8xdocs/epst/mod11/m11_010_000.html Link to image quality troubleshooting guide. Scan stops in the middle of gradient stressing (GRE) sequences. After the gradients recovered, the scan completes (eventually after several stops). The images get zippers and artifacts due to the stops.

Error message

2244588 The GP is reporting SGA Power Supply Power Off Fault. Dec 4 Tue 06:21:29 2244588 NSP SCP:Grad 739 The GP is reporting SGA Power Supply Power Off Fault.->

Applies to

Models
Excite
Product
Excite HD For 12.0 1.5T Systems only.
Software
12.0 Excite HD and 3T Excite HD products only
Area
MR