Unable to scan with DCERD Temperature Error 2266418 due to Equipment Room too Hot.
Symptom
Unable to scan with DCERD Temperature Error 2266418 See Long Description HDe system posts error 2266418, along with a string of errors that seem to point to a major problem in the System Cabinet. The errors mention SRF, WARP CPU, and B0 correction problems.
Error message
2244570 The gradient amplifier failed to go to Ready when commanded. 2244570 TOiMR2 NSP SCP:Grad 1106 Exception Class: Unknown Severity: Unknown Function: Gradient: N/A PSD: Spin Echo Coil: US SHOULDER PA Scan: 379/4 The X Axis gradient amplifier failed to go to Ready when commanded.-> 2244570 TOiMR2 NSP SCP:Grad 1106 Exception Class: Unknown Severity: Unknown Function: Gradient: N/A PSD: Spin Echo Coil: US SHOULDER PA Scan: 379/4 The X Axis gradient amplifier failed to go to Ready when commanded. 2244851 The GP is reporting a Gradient Amplifier %c Axis Current Distortion Fault. the same error as 2244577 on 11X, G3, openMR4 and 2244578 on 12X, E2, 14M4 (HDx, HDe), and openMR5. See Long Description 2244851 error code on Post 14M5 and 12.0 M5, is the same error as 2244577 on 11X, G3, openMR4and 2244578 on 12X, E2, 14M4 (HDx, HDe), and openMR5.2244851 The GP is reporting a Gradient Amplifier %c Axis Current Distortion Fault Example error message (could be any gradient axis, Y-axis is shown in this example): Error: 2244851 Host: lx-mr2 Process: NSP File: SCP:Grad Line: 2851 The GP is reporting a Gradient Amplifier Y Axis Current Distortion Fault.-> Extended error message This is a gen
Applies to
- Models
- Excite
- Product
- Excite HDe. For Hino HDe 14.0 and HDe 15.0 Software Systems only
- Software
- HDe 14.0 For Hino Excite HDe 14.0 systems only.
- Area
- MR
Full cause & step-by-step resolution
The complete cause analysis, the step-by-step remedy for this GE Healthcare MR issue are available to FSELIB subscribers — re-authored, web-first, searchable.
7-day free trial · no credit card required
Related GE Healthcare MR issues
- TPS/MGD fails to download or reset. mgd_term windows fail to mount file system.
- The Carriage Fails to Advance to Latch to the Table and to Retract When it is Raised, Undocked or Lowered.
- Unable to Load Applications Software during restoreinfo Due to Corrupt saveinfo Disk.
- CWA ID100315. Auto and Manual Prescan Fails.
- Ghosted or Smeared Head Images Using Head Coil due to Head TR Assembly.
- A Copper Coolant Line or Pipe Internal to the MCS Cabinet Chiller Burst due to Water Remaining in Unit Freezing During Shipment.