2247383 Software Issue APS detected memory corruption while creating output filename for the anti-alias test.

Symptom

RRF diags fail EX-RX signal test Error: 2247383 Software Issue: APS detected memory corruption while creating output filename for the anti-alias test.->

Error message

2247342 Master Exciter Anti-alias diagnostics failed. Please perform a TPS reset. Mon Jan 12 14:59:00 2009 Error: 2247342 Host: BRHDAGSLPMR Process: diagCoord File: rrfdpDiags.c Line: 675 Diagnostic: Ex-Rx Signal Iteration: 1 of 1 Action: 1 Task ID: 0x3a8d228 Time: 219222 Object: APS Test Task Master Exciter Anti-alias diagnostics failed. Please perform a TPS reset.

2247383 Software Issue APS detected memory corruption while creating output filename for the anti-alias test. Extended error: This error is the result of a coding error, Correct the filename size to the exciter or reciever response., possible cause software. Tried on another system , diags pass., Doesnt seem to be an issue, system scanning ok. From Software Engineering , Doesn't seem like it would be a hardware issue. Is it possible a bad or long name was entered into the diag? Itrack written.

Applies to

Models
Excite
Product
Excite Includes Excite 1 and Excite 2, 10.0, 11.0 and 11.1 Software.
Software
11.1 For product 1.5T Excite only. Note, 3T Excite goes with software G3, not 11.0 or 11.1.
Area
MR

Full cause & step-by-step resolution

The complete cause analysis, the step-by-step remedy for this GE Healthcare MR issue are available to FSELIB subscribers — re-authored, web-first, searchable.