GE Healthcare · MRExcite22473422192222247383
2247383 Software Issue APS detected memory corruption while creating output filename for the anti-alias test.
Symptom
RRF diags fail EX-RX signal test Error: 2247383 Software Issue: APS detected memory corruption while creating output filename for the anti-alias test.->
Error message
2247342 Master Exciter Anti-alias diagnostics failed. Please perform a TPS reset. Mon Jan 12 14:59:00 2009 Error: 2247342 Host: BRHDAGSLPMR Process: diagCoord File: rrfdpDiags.c Line: 675 Diagnostic: Ex-Rx Signal Iteration: 1 of 1 Action: 1 Task ID: 0x3a8d228 Time: 219222 Object: APS Test Task Master Exciter Anti-alias diagnostics failed. Please perform a TPS reset. 2247383 Software Issue APS detected memory corruption while creating output filename for the anti-alias test. Extended error: This error is the result of a coding error, Correct the filename size to the exciter or reciever response., possible cause software. Tried on another system , diags pass., Doesnt seem to be an issue, system scanning ok. From Software Engineering , Doesn't seem like it would be a hardware issue. Is it possible a bad or long name was entered into the diag? Itrack written.
Applies to
- Models
- Excite
- Product
- Excite Includes Excite 1 and Excite 2, 10.0, 11.0 and 11.1 Software.
- Software
- 11.1 For product 1.5T Excite only. Note, 3T Excite goes with software G3, not 11.0 or 11.1.
- Area
- MR
Full cause & step-by-step resolution
The complete cause analysis, the step-by-step remedy for this GE Healthcare MR issue are available to FSELIB subscribers — re-authored, web-first, searchable.
7-day free trial · no credit card required
Related GE Healthcare MR issues
- TPS/MGD fails to download or reset. mgd_term windows fail to mount file system.
- The Carriage Fails to Advance to Latch to the Table and to Retract When it is Raised, Undocked or Lowered.
- Unable to Load Applications Software during restoreinfo Due to Corrupt saveinfo Disk.
- CWA ID100315. Auto and Manual Prescan Fails.
- Ghosted or Smeared Head Images Using Head Coil due to Head TR Assembly.
- A Copper Coolant Line or Pipe Internal to the MCS Cabinet Chiller Burst due to Water Remaining in Unit Freezing During Shipment.