HDe TPS Reset Fails Due to VRE Configuration.

Symptom

During Application start-up, some or all of the Image Compute Nodes (ICN) may not successfully initiate and TPS Reset failures will occur See Resolution concept for the complete SMILe released 07/21/2006. Problem: Changing the host IP address or MGD IP address without performing an ICN Reconfiguration leaves the VRE configuration files with information pertaining to the old IP addresses. In this state, during Application start-up, some or all of the Image Compute Nodes (ICN) may not successfully initiate and TPS Reset failures will occur.

Applies to

Models
Excite
Product
Excite HDe. For Hino HDe 14.0 and HDe 15.0 Software Systems only
Software
HDe 14.0 For Hino Excite HDe 14.0 systems only.
Area
MR

Full cause & step-by-step resolution