Pular para o conteúdo principal
GE Healthcare · X-ray VascularInnova140001218000181400016

No Acquisition exposure, fluoro or record

Symptom

No Acquisition exposure, fluoro or record will not fluoro or record after boot. No issue with Positioner movement, DL, Rtac and detector is up. Innova 2100iq system will not fluoro or record after boot. No issue with Positioner movement, DL, Rtac and detector is up. Error 3200011

Error message

1400012: ERR_C80_PROC: AR: Component(s)_RTAC_C80_Boards: Async error - Card 2 (IP3) - C80_ERROR LEVEL 0 - Kernel: DSP error id: 21 ( param 0 ) 1400012: ERR_C80_PROC: AR: Component(s)_RTAC_C80_Boards: Async error - Card 2 (IP3) - C80_ERROR LEVEL 0 - Kernel: DSP error id: 21 ( param 0 ). 1800018 ERR SIB IP BOARDS LINKS IH XRExpSens DSP init Component(s) RTAC SIB C80 Boards links 1 0 [DSP automaticNaming] IP2 (card 1) -IP3 (card 2) link failure detected 1400016 ERR C80 MEZZANINE CRITICAL 1400012 ERR C80 PROC 1400012: ERR_C80_PROC: AR: Component(s)_RTAC_C80_Boards: Async error - Card 2 (IP3) - C80_ERROR LEVEL 0 - Kernel: DSP error id: 21 ( param 0 ) 1800018 ERR SIB IP BOARDS LINKS IH XRExpSens DSP init Component(s) RTAC SIB C80 Boards links 1 0 [DSP automaticNaming] IP2 (card 1) -IP3 (card 2) link failure detected 1400016 ERR C80 MEZZANINE CRITICAL 1400012 ERR C80 PROC

Applies to

Models
Innova
Product
Innova 2000 Innova 2000 systems Innova 2000 systems
Software
$, Innova 2000 Service Pack FMI F4, HerSvcPack_2.02 HerSvcPack_2.02 DL Applications version 11.23.26 Ghost DL-1 "F" : 17.4 Ghost DL-1 "G" : 15.5 Ghost DL-2 : 16.4 Ghost DL-4 : NA DL : 11.23.26 DL Atlas Core : 3.1O-1 RTAC Atlas Core : 3.1O-2 IDC : 2.13 Calypso : 11.13_FMI_7 Calypso Uptime : 11.25_FMI Titan : H0.57 Pos. Motorola : 5.2Y3-2 Pos. Roadrunner : 5.0R
Area
X-ray Vascular

Full cause & step-by-step resolution

The complete cause analysis, the step-by-step remedy for this GE Healthcare X-ray Vascular issue are available to FSELIB subscribers — re-authored, web-first, searchable.