The Advantx Console is Blank on boot of the system.

Symptom

The Advantx Console is Blank on boot of the system. The console green power light is on, but the system appears to have not booted and the console has no display. The Advantx Console is blank on boot of the system. The console green power light is on, but the system appears to have not booted and the console has no display.

Error message

IN2K: Ping Test from Digital Leader to Titan Kernal Fails Ping Test from Digital Leader to Titan Kernal Fails. The Ping from the Digital Leader (DL) to the Titan fails. Attempt to ping the Titan Kernal from the DL. Open a command window and Ping 165.100.10.3. If The system reports "request timed out", The Ping to the Titan Kernal has failed

IN2K: ECM has a PRD code of 72H on boot of the system. This indicates the ECM can not communicate with with the ASC or that the ASC is off line. Softstart will be off as a result. ECM has a PRD code of 72H on boot of the system. This indicates the ECM can not communicate with the ASC or that the ASC is off line. Softstart will be off as a result.

4000054 XRTitan connect call failed 4000054 XRTitan connect call failed: Component(s)_ETHERNET_Communication Root Cause : Component(s) - Ethernet Communication. This error is due to time out on Ethernet Communication with Titan subsystem.

4000048 could not send packet Component(s)_ETHERNET_Communication 4000048 : Recommended Actions: Root Cause : Component(s) - Ethernet Communication. This error is due to time out on Ethernet Communication. Recommended Actions: ping -t RTAC and Titan (165.100.10.2 an

Applies to

Models
Innova, Advantx
Product
Innova 2000 Innova 2000 systems Innova 2000 systems
Area
X-ray Vascular

Full cause & step-by-step resolution

The complete cause analysis, the step-by-step remedy for this GE Healthcare X-ray Vascular issue are available to FSELIB subscribers — re-authored, web-first, searchable.