GE Healthcare · MRExciteSigna2249152
T/R Driver Faults Due To SSM.
Symptom
Unable to Scan With T/R driver Fault, Body Path Short
Error message
2249152 T/R driver fault, Body Path short. Possibly a shorted diode in the body T/R switch network, or a short circuit. Fri May 21 16:15:34 2004 Error: 2249152 Host: lx-mr Process: NSP File: SCP:../TrDriverMdsHWO.cpp Line: 1310 T/R driver fault, Body Path short^A Possibly a shorted diode in the body T/R switch network, or a short circuit. Fri May 21 16:15:34 2004 Error: 2249152 Host: lx-mr Process: NSP File: SCP:../TrDriverMdsHWO.cpp Line: 1310 T/R driver fault, Body Path short. Possibly a shorted diode in the body T/R switch network, or a short circuit.
Applies to
- Models
- Excite, Signa
- Product
- Excite Includes Excite 1 and Excite 2, 10.0, 11.0 and 11.1 Software.
- Software
- 11.0 Only for 1.5T Excite product. 12.0 is Excite HD and 14.0 is Excite HDx. Note, 3T Excite goes with software G3, not 11.0 or 11.1.
- Component
- SecondOpinion System Support Module 2160200-43 See comments in resolution concept for second opinion requirements before ordering this part. FE replaced SSM which is now a FRU Part #: 2160200-43 Part Type: GP REPAIRABLES Part Weight: 48.50 LBR Repairable: Yes Returnable: Yes Hazardous Material: N Description: ERBTEC #550020 SYSTEM SUPPORT MODULE FRU: Yes Status: Orderable ICV/Transfer Price: 5684.00 USD /Each List Price: 36215.22 USD /EACH See comments in resolution concept for second opinion requirements before ordering this part. FE replaced SSM which is now a FRU Part #: 2160200-43 Part Type: GP REPAIRABLES Part Weight: 48.50 LBR Repairable: Yes Returnable: Yes Hazardous Material: N Description: ERBTEC #550020 SYSTEM SUPPORT MODULE FRU: Yes Status: Orderable ICV/Transfer Price: 5684.00 USD /Each List Price: 36215.22 USD /EACH
- Area
- MR
Full cause & step-by-step resolution
The complete cause analysis, the step-by-step remedy for this GE Healthcare MR issue are available to FSELIB subscribers — re-authored, web-first, searchable.
7-day free trial · no credit card required
Related GE Healthcare MR issues
- Running grafidy3, and find there are spikes on the waveforms in the manual prescan display, both windows.
- TPS/MGD fails to download or reset. mgd_term windows fail to mount file system.
- Fails to Scan with TORSO Coil.
- The Carriage Fails to Advance to Latch to the Table and to Retract When it is Raised, Undocked or Lowered.
- AE title has changed on destination host.
- Unable to Load Applications Software during restoreinfo Due to Corrupt saveinfo Disk.