Innova System Boot Failure, all monitors are blank on the Digital Leader.
Symptom
Innova System Boot Failure, all monitors are blank on the Digital Leader. The DL VGA monitor has a blinking light indicating no video. Power to monitors OK LED indicator on, but no display on DL flat panel, no image is present on the in room monitors. The system appears to not "turn on" for the customer Innova System Boot Failure, all monitors are blank on the Digital Leader. The DL VGA monitor has a blinking light indicating no video. Power to monitors OK LED indicator on, but no display on DL flat panel, no image is present on the in room monitors. The system appears to not "turn on" for the customer Innova System Boot Failure or all monitors are blank. DL did not boot. Sometimes on a second reset or boot the DL boots ok.
Error message
2800059 IH XRSubSysMgrDL Failure of XRImProc during boot of the application Debug: component startup timeout 2800059 IH XRSubSysMgrDL Failure of XRImProc during boot of the application Debug: component startup timeout. May be followed by error codes 2800060, 2800061, 2800066, 2800006 2800059 IH XRSubSysMgrDL Failure of XRImProc during boot of the application Debug: component startup timeout May be followed by error codes 2800060, 2800061, 2800066, 2800006 IN2K: 4F2D Xray aborted - Restart Xray; if problem persists, call Service Titan error 4F2D Xray aborted - Restart Xray; if problem persists, call Service
Applies to
- Models
- Innova
- Product
- Innova X100 Monoplane Innova X100 Monoplane from product name Innova 2100, 3100 and 4100 Innova X100 Monoplane Innova 2100, 3100 and 4100 Innova 2000 Innova 2000 systems Innova 2000 systems
- Component
- Digital Leader DL / RTAC Chassis power supply 2403760 or 2403760-2 RTAC/ DL POWER SUPPLY Digital Leader DL / RTAC Chassis power supply 2403760 or 2403760-2 RTAC/ DL POWER SUPPLY Digital Leader DL / RTAC Chassis power supply 2403760 or 2403760-2 RTAC/ DL POWER SUPPLY
- Area
- X-ray Vascular
Full cause & step-by-step resolution
The complete cause analysis, the step-by-step remedy for this GE Healthcare X-ray Vascular issue are available to FSELIB subscribers — re-authored, web-first, searchable.
7-day free trial · no credit card required
Related GE Healthcare X-ray Vascular issues
- The Worklist does not refresh on the system when performing a query.
- Innova System boot failure
- INX1X; Message Limit switch engaged; LP4 Lateral Detector lift will not move after going to max SID using the Emergency Backout button on the lift cover because the lift hit maximum (Max) SID limit sw…
- Firewall is broadcasting DHCP addresses
- Cooling Failure; Shut System down and call Service is displayed on the Innova Image monitor
- No gantry motion is reported by the customer.