Pular para o conteúdo principal
GE Healthcare · MRExcite2224842

Cradle will not move due to a broken coupling between motor and gearbox.

Symptom

Table Cradle will Move In Briefly then Stop See Long Description Cradle will not drive in completely.

Error message

2224842 SRI Hardware Error Unexpected cradle motion. Thu Jun 9 13:59:11 2005 Error: 2224842 Host: MR_PHC Process: NSP File: HPC_PORT_ID:sri Line: 99 SRI Hardware Error: Unexpected cradle motion. This error would occur if the table were moved by hand. Please run SRI diagnostics if that is not the case. Expected Cradle Velocity: -100 mm/sec Measured Cradle Velocity: 189 mm/sec

Applies to

Models
Excite
Product
Excite HD For 12.0 1.5T Systems only.
Software
12.0 Excite HD and 3T Excite HD products only
Component
46-282363P1 NON-MAGNETIC FLEXIBLE COUPLING
Area
MR

Full cause & step-by-step resolution

The complete cause analysis, the step-by-step remedy for this GE Healthcare MR issue are available to FSELIB subscribers — re-authored, web-first, searchable.

7-day free trial · no credit card required

Related GE Healthcare MR issues