GE Healthcare · MRExcite22542072254054
2254054 RF Amplifier in low power state avoid MEMP FSE scans Due to SRFD2 Amplifier Falsely States it is in Low Power St…
2254054 RF Amplifier in low power state avoid MEMP FSE scans Due to SRFD2 Amplifier Falsely States it is in Low Power State.
Symptom
System Still Scans With SRFD2 Error System is scanning but is getting SRFD2 low power errors in the log.
Error message
2254207 SRFD2 RF Amplifier Device Fault 44. See Long Description 2254207 RF FET device is shorted. SRFD2 will attempt to bypass this failure and run in low power mode. Replace Amplifier. 2254054 RF Amplifier in low power state avoid MEMP FSE scans Tue Oct 4 09:13:27 2005 Error: 2254054 Host: lx-mr Process: NSP File: HPC PORT ID:spi rf rd cs handler Line: 5 RF Amplifier in low power state; avoid MEMP, FSE scans.^A RF Amplifier has successfully bypassed a fault. IQ may be compromised. Please run prescan again. Scanning can then continue at reduced output levels. Scanning may be limited. Contact FE
Applies to
- Models
- Excite
- Product
- Excite HD For 12.0 1.5T Systems only.
- Software
- 12.0 Excite HD and 3T Excite HD products only
- Component
- SecondOpinion SRFD2 1.5T Solid State RF Amplifier 2351573 See comments in Resolution Concept for second opinion requirements before ordering this part. SecondOpinion SRFD2 1.5T Solid State RF Amplifier 2351573
- Area
- MR
Full cause & step-by-step resolution
The complete cause analysis, the step-by-step remedy for this GE Healthcare MR issue are available to FSELIB subscribers — re-authored, web-first, searchable.
7-day free trial · no credit card required
Related GE Healthcare MR issues
- TPS/MGD fails to download or reset. mgd_term windows fail to mount file system.
- The Carriage Fails to Advance to Latch to the Table and to Retract When it is Raised, Undocked or Lowered.
- Unable to Load Applications Software during restoreinfo Due to Corrupt saveinfo Disk.
- CWA ID100315. Auto and Manual Prescan Fails.
- Ghosted or Smeared Head Images Using Head Coil due to Head TR Assembly.
- A Copper Coolant Line or Pipe Internal to the MCS Cabinet Chiller Burst due to Water Remaining in Unit Freezing During Shipment.