GE Healthcare · MRExciteDiscovery

VRE Configuration Fails With System set to Non-English Language due to Software Bug.

Symptom

VRE Configuration Fails With System set to Non-English Language See Long Description iTrak # 13230392 15.0 m3 software: 15.0M3 software was being installed as an upgrade from 14.0M5a. As per instructions, operating system and application software had been loaded and RestoreInfo had been performed. The system had been configured to run in French with the 14x software so this setting had been imported when the RestoreInfo was done. VRE Configuration failed with no error messages, just a dialog box with an OK button and "VRE Configuration Failed." iTrak # 13224435 14.0 M5 software: WHEN THE USER INTERFACE LANGUAGE IS SET TO FRENCH WE ARE NOT ALLOWED TO CONFIGURE THE VRE or ICN

Error message

VRE Configuration Failed See Long Description VRE Configuration Failed VRE Configuration failed with no error messages, just a dialog box with an OK button and "VRE Configuration Failed."

Applies to

Models
Excite, Discovery
Product
Excite HDx Includes "Excite HDi" system. 14.0 Software only.
Software
14.0 M5 Excite HDx HDi Platforms only.
Area
MR

Full cause & step-by-step resolution