GE Healthcare · MRExcite222458022247252224666
Banana plug on the Head T/R switch has bent/broken fingers causing coil to disconnect.
Symptom
Can't get coil cable to stay connected
Error message
2224580 Auto Prescan failed. Signal was too small to process - use Manual Prescan 2224580 Host: hopgemr1a Process: NSP File: APS:aps_cf Line: 11 Auto Prescan failed. Signal was too small to process - use Manual Prescan 2224725 Low signal detected. Please check the signal level for receiver X (or receiver X while using Manual Prescan) See Long Description Error: 2224725 Host: hopgemr1a Process: NSP File: APS:aps_frame_range_check Line: 10 Low signal detected. Please check the signal level for receiver X (or receiver X while using Manual Prescan). 2224666 Auto Prescan failed, signal too small. Please try Manual Prescan. Fri Jan 23 14:39:55 2004 2224666 lx-mr psc PscMainControl.cc 1271 Auto Prescan failed, signal too small. Please try Manual Prescan.
Applies to
- Models
- Excite
- Product
- Excite HD For 12.0 1.5T Systems only.
- Software
- 12.0 Excite HD and 3T Excite HD products only
- Component
- 46-282460P1 STD, HEX BRASS UN INSULATED BANANA used on Multi coil plugs Banana Plugs guide pins used on Multicoil plugs 46-282460P1 Banana Plugs guide pins used on Multicoil plugs 46-282460P1
- Area
- MR
Full cause & step-by-step resolution
The complete cause analysis, the step-by-step remedy for this GE Healthcare MR issue are available to FSELIB subscribers — re-authored, web-first, searchable.
7-day free trial · no credit card required
Related GE Healthcare MR issues
- TPS/MGD fails to download or reset. mgd_term windows fail to mount file system.
- The Carriage Fails to Advance to Latch to the Table and to Retract When it is Raised, Undocked or Lowered.
- Unable to Load Applications Software during restoreinfo Due to Corrupt saveinfo Disk.
- CWA ID100315. Auto and Manual Prescan Fails.
- Ghosted or Smeared Head Images Using Head Coil due to Head TR Assembly.
- A Copper Coolant Line or Pipe Internal to the MCS Cabinet Chiller Burst due to Water Remaining in Unit Freezing During Shipment.