Pular para o conteúdo principal
GE Healthcare · MRExcite2262134

CoilID Fails Unless Cradle is at Home Position due to Intermittent Cable in Rear Pedestal Cable and Track Assembly.

Symptom

CoilID Fails Unless Cradle is at Home Position. Cannot Change Coils Unless Cradle is at Home Position See Long Description Cannot switch coils and get green coil ID light unless cradle is in home position. If cradle is not at home position, when plugging in a coil or switching coils, the coil ID light is red and the error shown below is logged, and cannot scan. Is there any requirement on HDe for the table to be at a certain position when changing coils? Tue Sep 2 07:14:13 2008 Error: 2262134 Host: hde Process: NSP File: SCP:SRI Line: 348 Exception Class: Unknown Severity: Unknown Function: Patient Handling: Control/Display PSD: Unknown Coil: Unknown Scan: -1/-1 Coil validation error: Coil ID device was not detected on a coil. Coil ID is required for this connector. Please try connecting coil again.

Error message

2262134 Coil validation error. Coil ID device was not detected on a coil. Coil ID is required for this connector See Long Description Tue Sep 2 07:14:13 2008 Error: 2262134 Host: hde Process: NSP File: SCP:SRI Line: 348 Exception Class: Unknown Severity: Unknown Function: Patient Handling: Control/Display PSD: Unknown Coil: Unknown Scan: -1/-1 Coil validation error: Coil ID device was not detected on a coil. Coil ID is required for this connector. Please try connecting coil again. Extended Error: This could be a problem with the coil ID communication chain which includes the Coil, the cable between the Coil and the LPCA, the LPCA, the cable between the LPCA and SRI, and the SRI. Please run Coil ID diagnostics.

Applies to

Models
Excite
Product
Excite HDe. For Hino HDe 14.0 and HDe 15.0 Software Systems only
Software
HDe 14.0 For Hino Excite HDe 14.0 systems only.
Component
5151984 VALUE, REAR PEDESTAL, CABLE AND TRACK ASSY
Area
MR

Full cause & step-by-step resolution

The complete cause analysis, the step-by-step remedy for this GE Healthcare MR issue are available to FSELIB subscribers — re-authored, web-first, searchable.

7-day free trial · no credit card required

Related GE Healthcare MR issues