GE Healthcare · MRExcite22493721706269
Cannot Boot Applications with Host Monitor Timeout Error due to YP(NIS) Service was turned "On" in Network Configuration…
Cannot Boot Applications with Host Monitor Timeout Error due to YP(NIS) Service was turned "On" in Network Configuration in Guided Install.
Symptom
Cannot Boot Applications with Host Monitor Timeout Error
Error message
2249372 Communication problem with the host. Remote command execution (rcmd) failed. See long description for complete error. Fri Apr 1 16:52:38 2005 Error: 2249372 Host: lx-mr Process: diagCoord File: mgdPosixSync.c Line: 1215 Task ID: 0x3ff00c0 Time: 1706269 Object: APS Communication problem with the host. Remote command execution (rcmd) failed. Host: host; RSH Port: 514; Local User Name: genesis Remote User Name: genesis; Remote Command: getHostTime; fd2p pointer: 0->
Applies to
- Models
- Excite
- Product
- Excite Includes Excite 1 and Excite 2, 10.0, 11.0 and 11.1 Software.
- Software
- 11.1 For product 1.5T Excite only. Note, 3T Excite goes with software G3, not 11.0 or 11.1.
- Area
- MR
Full cause & step-by-step resolution
The complete cause analysis, the step-by-step remedy for this GE Healthcare MR issue are available to FSELIB subscribers — re-authored, web-first, searchable.
7-day free trial · no credit card required
Related GE Healthcare MR issues
- TPS/MGD fails to download or reset. mgd_term windows fail to mount file system.
- The Carriage Fails to Advance to Latch to the Table and to Retract When it is Raised, Undocked or Lowered.
- Unable to Load Applications Software during restoreinfo Due to Corrupt saveinfo Disk.
- CWA ID100315. Auto and Manual Prescan Fails.
- Ghosted or Smeared Head Images Using Head Coil due to Head TR Assembly.
- A Copper Coolant Line or Pipe Internal to the MCS Cabinet Chiller Burst due to Water Remaining in Unit Freezing During Shipment.