GE Healthcare · MR
OrthoMRI Blown fuse in the SRFA module OrthoMRI Unable to scan due to RFA failure Replaced SRFA fuse OrthoMRI Escan XQ O…
OrthoMRI Blown fuse in the SRFA module OrthoMRI Unable to scan due to RFA failure Replaced SRFA fuse OrthoMRI Escan XQ OrthoMRI SRFA OFF Direct power supply failure to the RF amplifier
Symptom
OrthoMRI Unable to scan due to RFA failure
Error message
OrthoMRI SRFA OFF Direct power supply failure to the RF amplifier This event may also be due (as well to the fuse being blown, or disconnected 220 V AC cable etc.) to overheating of the amplifier (T > 50 degrees C)
Applies to
- Product
- OrthoMRI Escan XQ EscanXQ and EscanNE
- Software
- OrthoMRI S/W Release 9.1B. OrthoMRI S/W Release 9.1B.
- Component
- OrthoMRI LNR9102032000 FUSE SET If any blown fuse is not available in the fuse kit provided with the system order the Fuse Set. If any blown fuse is not available in the fuse kit provided with the system order the Fuse Set. LNR9102032000 FUSE SET
- Area
- MR
Full cause & step-by-step resolution