OrthoMRI Corroded/Dirty network printer cable connector OrthoMRI Open association error when sending 3 on 1 images to th…

OrthoMRI Corroded/Dirty network printer cable connector OrthoMRI Open association error when sending 3 on 1 images to the printer OrthoMRI Cleaned network printer cable connectors OrthoMRI Escan XQ Or…

Symptom

OrthoMRI Open association error when sending 3 on 1 images to the printer No errors when sending to remote hosts nor when sending 1 on 1 images to printer.

Error message

OrthoMRI Open Association Error This error can be from a disconnected cable or the MRI system's IP address is not present or incorrect and lastly, the remote host is not set up properly. OrthoMRI Open Association Error

OrthoMRI Send or Receive failure error message

Applies to

Product
OrthoMRI Escan XQ EscanXQ and EscanNE
Software
OrthoMRI S/W Release 9.1B. OrthoMRI S/W Release 9.1B.
Area
MR

Full cause & step-by-step resolution