GE Healthcare · MRExcite2258107
2258107 Coil Check Host coil validation error Body Transmit Enable mismatch detected Due to large cable track and harnes…
2258107 Coil Check Host coil validation error Body Transmit Enable mismatch detected Due to large cable track and harness.
Symptom
Any receive only coil connected to Port A immediately generates error 2258107 Coil ID light goes through normal sequence (brief red, then green) then error 2258107 is logged.
Error message
2258107 Host coil validation error Body Transmit Enable mismatch detected Error: 2258107 Host: MR3T Process: coilCK File: coilCK scp.c Line: 1026 Exception Class: Unknown Function: Host: Coil Check Host coil validation error: Body Transmit Enable mismatch detected. Connector: Multicoil A Port Coil: HD Shoulder Array by GE - A/B port
Applies to
- Models
- Excite
- Product
- Excite HDx Includes "Excite HDi" system. 14.0 Software only.
- Software
- 14.0 M5 Excite HDx HDi Platforms only.
- Component
- 5115097-2 CABLE AND TRACK LARGE LONG ASM Carriage Ports to 16 Channel Switch 5115097-2 CABLE AND TRACK LARGE LONG ASM Carriage Ports to 16 Channel Switch
- Area
- MR
Full cause & step-by-step resolution
The complete cause analysis, the step-by-step remedy for this GE Healthcare MR issue are available to FSELIB subscribers — re-authored, web-first, searchable.
7-day free trial · no credit card required
Related GE Healthcare MR issues
- TPS/MGD fails to download or reset. mgd_term windows fail to mount file system.
- The Carriage Fails to Advance to Latch to the Table and to Retract When it is Raised, Undocked or Lowered.
- Unable to Load Applications Software during restoreinfo Due to Corrupt saveinfo Disk.
- CWA ID100315. Auto and Manual Prescan Fails.
- Ghosted or Smeared Head Images Using Head Coil due to Head TR Assembly.
- A Copper Coolant Line or Pipe Internal to the MCS Cabinet Chiller Burst due to Water Remaining in Unit Freezing During Shipment.