Pular para o conteúdo principal
GE Healthcare · MRExcite13214745

The Guided Install Configuration Language Was Improperly Set.

Symptom

Guided Install Displayed in Foreign Language See Long Description An FE reported that during the PM he launched the Guided Install GUI in FE Mode and found all tab labels and other associated text on the GUI displayed in Japanese instead of English. The config entries in the Guided Install text fields were all in English. All other text on the system was displayed in English as expected, including on the service browser and on the applications screens, only the text displayed in Guided Install was affected. When he checked the [Language Setting] under [System Configure] in Guided Install - FE Mode it was set to English. The FE rebooted the system with no effect on the problem. GI was still displayed in this foreign language and the [Language Setting] under [System Configure] in Guided Install - FE Mode was still set to English. iTrak # 13214745 .

Applies to

Models
Excite
Product
Excite HDx Includes "Excite HDi" system. 14.0 Software only.
Software
14.0 M5 Excite HDx HDi Platforms only.
Area
MR

Full cause & step-by-step resolution

The complete cause analysis, the step-by-step remedy for this GE Healthcare MR issue are available to FSELIB subscribers — re-authored, web-first, searchable.

7-day free trial · no credit card required

Related GE Healthcare MR issues