No Documented Procedure to Perform checkINFO on HFO4 Systems. checkINFO Procedure for HFO4 Systems.

Symptom

No Documented Procedure to Perform checkINFO on HFO4 Systems See Long Description iTrak # 13214413 There is no procedure included in Signa Excite OpenSpeed Service Methods, Revision 9 (latest version) to perform checkINFO on INFO MODs on HFO4 systems. In the document "Before Starting a Load From Cold", section 5.1 "SaveInfo", step 5 states as follows: "When SaveInfo is complete, click [Verify SaveInfo Media] and make sure all information is saved correctly and verify the SaveInfo date." But this button [Verify SaveInfo Media] only exists on HFO5 systems . It is absent on HFO4 systems. Therefore no documented procedure exists for HFO4 systems. How do we perform this function on HFO4 systems? To view the most recent appropriate and detailed documentation, click to go to MR Service Documentation.

Applies to

Models
Signa, Excite
Product
Open Speed with EXCITE For HFO4 and HFO5 only
Software
OPENMR4-HFO4 Software version for HFO4 system, which is OpenSpeed with Excite Technology
Area
MR

Full cause & step-by-step resolution