High Speed Stability 3.0T (HSS) Procedure Lists a Non-3.0T HSS Kit 46-320383G1 as Tools and Test Equipment.
Symptom
High Speed Stability 3.0T (HSS) Procedure Lists a Non-3.0T HSS Kit 46-320383G1 as Tools and Test Equipment See Long Description In the 3.0T HSS procedure (High Speed Stability 3.0T) as found in Signa EXCITE HD 3.0T Service Methods, Dir 5133330-1, rev 10, under step 3.1 "Tools and Test Equipment", the following tool is listed for performance of this procedure: UNIVERSAL SST KIT 3.0T 46-320383G1 The problem is, this is not a 3.0T SST kit. This is actually a "UNIVERSAL SMALL SAMPLE TEST KIT FOR SIGNA 1.5/0.5T AND MR MAX " and can't be used on 3T. iTrak # 13222921 To view the most recent appropriate and detailed documentation, click to go to MR Service Documentation.
Applies to
- Models
- Excite, Signa
- Product
- 3T Excite HD 12.0 software only
- Software
- 12.0 Excite HD and 3T Excite HD products only
- Area
- MR