GE Healthcare · MR22550872255085
2255087 Shim Supply ADC Gain Error due to failed Z2 Supply.
Symptom
Auto prescan failed with Shim power supply errors Auto prescan failed with Shim power supply errors.
Error message
2255087 Shim Supply ADC Gain Error. Detected Reference = 0 Counts. 2255087 IC MR2 NSP SCP:Shim 434 Shim Supply ADC Gain Error. Detected Reference = 0 Counts. 2255085 Internal hardware error in the Shim Supply: Low Rail Voltage. Positive Rail = XX mVolts Negative Rail = XX mVolts. 2255085 2 IC MR2 NSP SCP:Shim 829 Internal hardware error in the Shim Supply: Low Rail Voltage. Positive Rail = XX mVolts Negative Rail = XX mVolts.-
Applies to
- Product
- Open Speed with EXCITE For HFO4 and HFO5 only
- Software
- OPENMR5-HFO5 Open Speed with Excite Product
- Component
- 5106397 5 Channel Shim Power Supply
- Area
- MR
Full cause & step-by-step resolution
The complete cause analysis, the step-by-step remedy for this GE Healthcare MR issue are available to FSELIB subscribers — re-authored, web-first, searchable.
7-day free trial · no credit card required
Related GE Healthcare MR issues
- Running grafidy3, and find there are spikes on the waveforms in the manual prescan display, both windows.
- OrthoMRI Network cable is unplugged
- TPS/MGD fails to download or reset. mgd_term windows fail to mount file system.
- CWA ID100060 Unable To Film.
- Fails to Scan with TORSO Coil.
- The Carriage Fails to Advance to Latch to the Table and to Retract When it is Raised, Undocked or Lowered.