GE Healthcare · MR2254278
CWA ID119427. Will Not Scan With RF Errors.
Symptom
Will Not Scan With RF Errors Will not scan with RF errors
Error message
2254278 3T RF Amplifier Enable Line fault 93. The RF amplifier indicates that it was unblanked continuously for an excessive amount of time. Error: 2254278 Host: CENTMR02 Process: NSP File: SCP:../rf_nb_contro Line: 1491 3T RF Amplifier Enable Line (fault 93). The RF amplifier indicates that it was unblanked continuously for an excessive amount of time. Check PSD parameters. If the fault persists, replace the RF amplifier.
Applies to
- Product
- 3T Eclipse For E2 software
- Software
- G3 For 3T Excite only
- Area
- MR
Full cause & step-by-step resolution
The complete cause analysis, the step-by-step remedy for this GE Healthcare MR issue are available to FSELIB subscribers — re-authored, web-first, searchable.
7-day free trial · no credit card required
Related GE Healthcare MR issues
- Unable to scan with RF Error 2254278 due to SSM.
- 2254278 3T RF Amplifier Enable Line fault 93 Due To Defective RF Amplifier.
- Will Not Scan With RF Errors. 2254278 3T RF Amplifier Enable Line fault 93.